is designed to measure the apparent resistivity with simultaneous forced probe configuration switching over (lateral sonde-normal sonde) and to measure rock spontaneous potential.
The measurements are performed via environmental excitation by alternating electric field (electrodes A, B) and potential drop measurement (at M, N electrodes). The spontaneous potential is measured when the external electric field is turned off.
KSP-43 Electrical logging tool is to be run with one-core and triple-core logging.
Sonde formula: normal sonde lateral sonde |
N1,0М0,1A А1,0М0,1N |
Apparent resistivity measurements range, Ohm·m |
0 ÷ 50000 |
SP measurement range, mV |
±10000 |
Basic relative measurements error, % |
up to 5 |
Supply voltage, V |
40-50 |
Output measure point (between N and M), mm: normal sonde lateral sonde |
195 498 |
Max. hydrostatic pressure, MPa |
20 |
Max. operating temperature, °С |
80 |
Tool dimensions, mm: diameter length (without a probe) length (including a probe) |
43 60 2700 |
Weight, kg |
15 |